    User name Yi Qiang

    Log entry time 13:30:53 on November 20, 2008

    Entry number 249977

    Followups:

    keyword=NMR Frequency/Field Sweep Cross Calibration for Vertical Pumping

    A Frequency/Field Sweep Cross Calibration was performed for vertical
    pumping. However, a glitch occurred before we rotated the field back to
    vertical direction. Therefore the second NMR frequency sweep results has
    quite large polarization loss compared to the first one and not reliable.
    We will try to another one at a later time.

    12:21 NMR Frequency Sweep@He3, Vertical+ Field, NMR Amplitude:
    28.17(27.54)mV -> Polarization: 68.31(66.78)%

    12:46 NMR Field Sweep@Pickup Coil, Longitudinal+ Field, NMR Amplitude:
    286.5+284 mV -> Polarization: 69.60%

    12:48 A small glitch occurred.

    13:11 NMR Frequency Sweep@He3, Vertical+ Field, NMR Amplitude:
    24.95(24.33)mV -> Polarization: 60.50(59.00)%




    Figure 1



    Figure 2



    Figure 3