    User name Chiranjib

    Log entry time 01:08:27 on November 21, 2008

    Entry number 250118

    keyword=Internal Temperature test

    Yi and I performed a test yesterday. Its just a late entry.

    Temperature Test on Cell ASTRAL for vertical pumping

    Laser StatusNMR signal(mV)RTD P1RTD P2RTD1RTD2RTD3RTD4RTD5
    ON21.47247.676246.2738.632871.1133137.59860.966837.5781
    OFF20.95224.004220.54738.6523134.73660.888760.888737.7051
    ON20.57248.437247.55938.62369.7559140.0261.591838.291



    COMMENTS
    The test is performed with a WAIT TIME of 20 min between two successive NMR(4 AFPs) measurements.RTDP1 and RTDP2 refer to the two RTD readouts from the pumping chamber surface.I will follow up with the temperature calculation closely later. But from my small simulation earlier, its seen that even with a difference of 50 degree C in the pumping chamber, the change in the polarization is not more than 2%.So our present polarization number with this temperature is good enough for the time being to make an estimation already as far as the temperature dependence is concerned. Of course here I assume that the target chamber temperature ( " the average target chamber temperature") stays almost constant and it is actually constant.