Log entry time 01:08:27 on November 21, 2008
Entry number 250118
keyword=Internal Temperature test
Yi and I performed a test yesterday. Its just a late entry.
Temperature Test on Cell ASTRAL for vertical pumping
Laser Status | NMR
signal(mV) | RTD P1 | RTD
P2 | RTD1 | RTD2 | RTD3 | RTD4 | RTD5 |
ON | 21.47 | 247.676 | 246.27 | 38.6328 | 71.1133 | 137.598 | 60.9668 | 37.5781 |
OFF | 20.95 | 224.004 | 220.547 | 38.6523 | 134.736 | 60.8887 | 60.8887 | 37.7051 |
ON | 20.57 | 248.437 | 247.559 | 38.623 | 69.7559 | 140.02 | 61.5918 | 38.291 |
COMMENTS
The test is performed
with a WAIT TIME of 20 min between two successive NMR(4 AFPs)
measurements.RTDP1 and RTDP2 refer to the two RTD readouts from the
pumping chamber surface.I will follow up with the temperature
calculation closely later. But from my small simulation earlier, its
seen that even with a difference of 50 degree C in the pumping
chamber, the change in the polarization is not more than 2%.So our
present polarization number with this temperature is good enough for
the time being to make an estimation
already as far as the temperature dependence is concerned. Of course
here I assume that the target chamber temperature ( " the average
target chamber temperature") stays almost constant and it is actually
constant.