User name V. Sulkosky
Log entry time 11:01:38 on November 21, 2008
Entry number 250175
This entry is a followup to: 250159
keyword=Re: U10 BCM Double Pulsing (fixed)
I found that the source of the second pulse for the BCM U10 signal was the
level translator in the middle room of the Counting House. This LT is
used to send a copy of the BCM signals to each spectrometer (left and
right). The NIM output going to the left HRS had two pulses.
After consulting with Bryan, I took the TTL output for channel 13 of the LT
and connected it to the input of channel 16. I then took one of these
outputs for the left HRS. These outputs are clean and only have one pulse.
The scaler rates for U10 for the left HRS and BigBite are now comparable to
the D10 signal. This work was completed during run 3787.
The first good run with this fix will be 3788.