    keyword=BigBite Gas Cerenkov Study II ----- ADC

    Following JP's suggestion, we did a study on ADC only. Here, we have the
    correlation cut. We did not apply the TDC cut, since it is biased. See
    fig. 1, the red one is "pion" and the black one is "electron".  The
    sample is reasonable. 
    

    Then we applied a cut at 200 to see the pion rejection and electron efficiency. From this study and my previous log entry, we can see clearly that the Gas Cerenkov somehow still miss a lof of electrons. The pion signal in ADC is due to background (not clear what). The pion does not fire the cerenkov, but the random background will.

    number of electron, electron eff, number of pions, pion percentage left within cut.

    869 0.168009 4350 0.0643678

    2287 0.390905 10566 0.308726

    2600 0.327692 2452 0.257341

    3711 0.156292 2243 0.0994204

    7017 0.142654 3806 0.0987914

    6481 0.13825 3575 0.067972

    30270 0.104063 25872 0.0903293

    5277 0.111806 2500 0.0788

    30270 0.115857 25872 0.104205

    600 0.223333 179 0.0893855

    273 0.216117 1344 0.00818452

    1400 0.254286 5067 0.00414446

    1833 0.270049 1597 0.0125235

    2177 0.205788 1350 0.00296296

    3067 0.0798826 1981 0.00504796

    3194 0.0491547 2219 0.0144209

    15293 0.000196168 14277 0.000560342

    2460 0.331707 1574 0.160737

    15293 0.147715 14277 0.134902

    282 0.180851 89 0.0337079