Log entry time 10:46:37 on November 26, 2008
Entry number 250883
keyword=EPR Calibration for Maureen of Vertical Pumping
Did two sets of EPR test of Maureen on Vertical Pumping. First one is
done with parallel pumping and second one is done with anti-parallel pumping:
Cell: Maureen
Oven Temperature: 230C
Laser Power: 75 W(Comet 25Wx3)
Target Position: He3
Field Direction: Vertical +
NMR RF Field: Longitudinal
NMR Pumping Chamber Pickup Coil: Transverse
Parallel Pumping: 5.23 %/mV
Anti-Parallel Pumping: 5.24 %/mV
Max polarization we got after yesterday's pumping is estimated to be 70%