    User name Moffit

    Log entry time 12:59:17 on November 28, 2008

    Entry number 251184

    keyword=HAPPEX DAQ: investigation of dx3 signal to happex7 scaler.

    Took another look at the dx3 signal during the Moller measurement. (Eugene uses the dx10 signal, so I dont think we were messing up his measurement).

    Injected a voltage (randomly changed each helicity window from the HAPPEX DAQ timing board dac12 output) into the v2f handling the signal from the dx3 downconverter. Using this, we were able to see that there were periodic double pulses into the cable going to the happex7 scaler.

    Found that a copy of this signal was sent to a cable running to the 2nd floor cage. This cable was unterminated. So I unplugged this cable at the Level Translator. Maybe someone was using this signal (at some point) to help commission the GEM detector in the hall.

    Signal to HAPPEX daq looks fine now... but we will continue to monitor it.