User name E.Chudakov
Log entry time 15:43:53 on November 28, 2008
Entry number 251193
keyword=Moller results
Moller results on Nov 28 12:50-14:40
- The beam tune took 2h. The main problem was tripping the EP ion
chamber. The beam strongly zig-zagged in Y (see
entry
251185).
- The energy lock on Hall C, pulsed, was used. At 0.44 µA in
Hall A the "HALLA:p" energy measurement is not very reliable and
depends on the beam current (see the strip chart). On the other
hand, the ARC-2 measurement "ARC2:p" is stable. The regular value
of "HALLA:p" obtained at higher current with the Hall A energy lock
is 5892 MeV.
- The difference with the previous measurements (about 74%) comes from the
beam energy difference.
Preliminary results: not corrected for
variations along the targets
# |
beam |
runs |
target |
Ptarget% |
Pbeam% |
1 | 0.44 µA |
12971-12974 | 5 | 7.97 |
81.6±0.4±2.0 |
2 | 0.44 µA |
12975-12980 | 4 | 7.44 |
82.1±0.3±2.0 |
3 | 0.44 µA |
12981-12984 | 3 | 7.62 |
77.6±0.3±2.0 |
4 | 0.44 µA |
12985-12988 | 2 | 8.21 |
80.8±0.2±2.0 |
5 | 0.44 µA |
12989-12990 | 5 | 7.97 |
81.6±0.5±2.0 |
6 | 0.62 µA |
12991-12994 | 3 | 7.62 |
78.0±0.3±2.0 |
7 | 0.62 µA |
12995-12996 | 5 | 7.97 |
80.4±0.4±2.0 |
average 1-4 | 80.5±0.2±2.0
|
A copy of this log entry has been emailed to: glamazdi@jlab.org