    User name Yi Qiang

    Log entry time 08:55:54 on December 02, 2008

    Entry number 251717

    Followups:

    keyword=EPR Measurement before Hall A access

    Cell: Maureen Oven Temperature: 230C Laser Power: 75 W(Comet 25Wx3) Target Position: He3 Field Direction: Transverse - NMR RF Field: Longitudinal NMR Pumping Chamber Pickup Coil: Vertical Pumping Direction: Anti-Parallel

    First Measurement: Integration Gain: 1.6 Calibration Constant: 2.95 %/mV Polarization: 62.6%

    Second Measurement: Integration Gain: 1.0 Calibration Constant: 2.92 %/mV Polarization: 59.7%

    The second measurement clearly has a better lock and accuracy, so I put that number to be our new calibration constant

    A copy of this log entry has been emailed to: jpchen,chiran,jinhuang