    User name R. Michaels

    Log entry time 12:03:39 on December 5,2008

    Entry number 252149

    This entry is a followup to: 252147

    keyword=polarity flip confirmation

    First of all, Ed Folts stopped by and confirmed that the
    L-HRS is ready. The magnets are ramped up to P0 = 2.35.
    Second, the tools GUI shows "P" for positive. See fig 1.
    Third, this has been known to be false in the past, so we
    checked the gaussmeters and dipole NMR :

    Gaussmeter .... before flip ..... after flip (now)

    Q1 .......... -0.15 ......... +0.15
    Q2 .......... +0.53 ......... -0.53
    Q3 .......... -0.47 ......... +0.47

    NMR
    D1 .......... 0.87 .......... 0.87 "positive"

    Waiting for beam ...