• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name R. Michaels

    Log entry time 12:55:45 on December 5,2008

    Entry number 252165

    keyword=re-establishing beam, return to production

    12:18 Beam returns, 2 uA
    Did BeO check.
    Checked BPMA, BPMB --> ok within tolerance
    Checked Spot++ --> ok, see fig 1.
    Put in He target.
    Do the raster / rate scan (2x2, 3x4, 4x5, 5x6.6)
    Run #4222 during the raster scan.
    Raster ... T1-rate ... T3-rate
    3x4 .... 13.5 kHz .... 480 Hz
    2x2 .... 14 kHz ...... 460 Hz ... event 9000
    4x5 .... 13 kHz ...... 480 Hz ... 23K
    5x6.6 .. 13.5 kHz .... 480 Hz ... 31K

    Rates pretty stable -- see fig 2.

    Returning to 12 uA production

    We do the following for Helium target:

    1. Manual flip ("Flip Now" button)
    2. Set auto flip on.

    Start run 4223, first production.



    FIGURE 1

    FIGURE 2