keyword=Swapped GSO crystal back into Compton
During this morning's access, Gregg Franklin and I visually inspected the GSO crystal and found it to be uncracked and clear. We replaced the PMT and tube base (for a tube base that draws much less current) and put the GSO crystal back in. Currently it's the GSO crystal that's in the photon beamline and connected to both DAQs. We hope to run it at -1700 V.