    Log entry time 13:53:16 on December 16, 2008

    Entry number 253623

    keyword=BB Cerenkov modifications

    PMTs 4, 5, 6, 7 on the beamline side of the Cerenkov are now immediately run through a Philips (NIM) 10x amplifier mounted right on the Cerenkov tank. The signals then run through the 10x amplifier on the front-end rack as usual. HVs on those PMTs were dropped to offset the extra amplification. LED runs were taken to check the single p.e. peak position and pedestal widths. (I'll post a few snapshots later.)

    Bottom line is that the resolution is definitely degraded, but not too badly. Hopefully this will let me run those PMTs safely under the present conditions.

    The TDC signal for PMT 7 has been missing for at least a week. I tracked it down to a bad (half of a) summing module. I moved the signals to the unused half of the same module and it seems to be OK now.

    I remeasured the discriminator threshold on the per-PMT discriminators and it was (still) -100mV.