    User name Jack

    Log entry time 14:11:08 on January 05, 2009

    Entry number 254878

    This entry is a followup to: 254873

    keyword=BigBite gas system status

    Xin started the gas flowing around 10AM. The ethanol bubbler was enabled, so the gas is flowing through the ethanol. Over the down I found a few leaks in the gas system and repaired them. The system, with the bubbler bypassed, has significantly less than 0.1 SCFH leak rate with 12 PSI pressure. With the bubbler in use the leak rate was not measured directly, but less than 0.1 SCFH more than the wire chambers were set to. So, we can expect to have reduced consumption of the premixed gas.