keyword=BigBite gas system status
Xin started the gas flowing around 10AM. The ethanol bubbler was
enabled, so the gas is flowing through the ethanol. Over the down I
found a few leaks in the gas system and repaired them. The system, with
the bubbler bypassed, has significantly less than 0.1 SCFH leak rate with
12 PSI pressure. With the bubbler in use the leak rate was not measured
directly, but less than 0.1 SCFH more than the wire chambers were set to.
So, we can expect to have reduced consumption of the premixed gas.