    User name Yi/Chiranjib/Jin

    Log entry time 00:42:56 on January 10, 2009

    Entry number 255276

    keyword=Transverse pumping chamber pickup coil adjustment

    During the final check of the target system before closing the target
    chamber, we found out that the transerse pumping chamber pickup coil
    inside the oven gave only 3 Ohm resistance with 230C oven temperature,
    compared to 28 Ohm of the pairing coil outside the oven. Looks like
    there's something going wrong of the coil, maybe short circuited.
    However, as we tested this before with the room temperature, the coil was
    normal. So we decided to shut down the heater. As the oven temperature
    dropping, we clearly saw the resistance of the inside coil was increasing
    and finally became 25 Ohm which is normal.

    At room temperature we rotated the field to vertical and tried to do NMR
    measurement with single pickup coil. The signal amplitude from only the
    outside coil itself is very consistant with the polarization we measured
    in the transverse direction using vertical pickup coils considering the
    ratio of the calibration constants we previously got from the two pumping
    directions. And the inside coil (even with 25 Ohm) gives a magnitude
    smaller signal!! This implies one thing, the inside coil was not working
    from the very beginning of the experiment!!!!! (Actually, the factor of
    two difference in the calibration constants was alreadly trying to tell
    us this, and unfortunately we just ignored that :S)

    Then we turned on the longitudinal RF and checked the RF signal pickuped
    by the two transverse coil. Inside coil gives 40 mV and outside coil
    gives 120 mV. (The orginal ajustment was done with water cell) Again, we
    turned on the heater, the signal from inside dropped to 5 mV and outside
    stays the same, 120 mV.

    Since the inside coil has a wierd problem and was not really used in the
    experiment, we decided to disconnect it. And did some adjustment of the
    outside coil to lower the RF background to 40 mV. After this, we clearly
    saw some improvement in the NMR measurement in terms of signal to
    background ratio.

    As a conclusion, we were using single pickup coil for vertical pumping
    all the time and may continue to use like this since replacing the inside
    coil will involve a lot of work including taking the oven apart and most
    importantly, the single coil worked already! Furthmore, we should have
    better signal to background ratio from now on.

    A copy of this log entry has been emailed to: jpchen