    User name Chiranjib

    Log entry time 15:34:23 on January 11, 2009

    Entry number 255463

    keyword=EPR before the Internal temp. test

    Holding field = Transverse -
    Cell = Brady
    Oven Temp. = 230 C

    Polarization = 58.39 %
    Calib constant = 2.79 %/mV

    Since I dont want to destroy too much polarization before the internal
    temp. test, I did not optimize the EPR AFP . I already destroyed some by
    doing few AFPs as it was difficult to lock in first couple of tries.

    DATA ::: C:\Happy Valentine\EPR\Data\EPR_AFP_20090111_1507



    NOTE ::: One of the RTDs on the pumping chamber was reading 302 C which
    was corrected in the analysis by following the old temp. readings. So the
    temp. 260 in the analysis is the average of the two RTDs. One of the RTDs
    was at 254 C and the other one was at 266 C ( corrected)

    A copy of this log entry has been emailed to: jpchen,yqiang



    Figure 1: EPR RF



    Figure 2 : EPR AFP



    Figure 3 : ANALYSIS



    Figure 4 : Just a freq. sweep NMR before the EPR