    User name Jin

    Log entry time 19:45:57 on January 11, 2009

    Entry number 255483

    This entry is a followup to: 255301

    Followups:

    keyword=Target Tests / Measurement

    
    Couple measurement Has been performed on target this evening:
    
    1. Trans. Pick Up Coil Performance Test:
       Signal Noise Ratial of Transverse Coils is 50% +- 5% of that of 
    Vertical ones. (20% noisier and 40% less sensitive)
    
    2. A NMR Cross Calibration is performed
    
    3. Spin Flip Using Field Rotation is tested to be working great
       Polarization Loss Rate of Spin Flip Using Field Rotation is < 200ppm
       Currently take 1s to flip. Could finish flip on the level of 
    milliseconds.
       Should not be sensitive to field gradiant
       
    4. Top Polarization of Last Spin up is 65.7%, spin up time = 3.82h
       Started at Halog 255301
       http://www.jlab.org/~adaq/halog/html/0901_archive/090110161506.html
    
    
    --------------------- Time Line & Details --------------------- 
    
    
    Start: 
    	Cell Brady, Oven @ 230C
    	3x Comet Laser pumping and several test after that
    	5x amplication for all preamplifier
    	
    
    16:14  Laser OFF
    
    ===================================
          Trans. Pick Up Coil Performance Test
    ===================================
    16:37  NMR Freq Sweep
    	Data ID : NMR_Freq_He_20090111_163727
    	-18.02 +- 0.07 mV for Sweep Up
    	18.3 +- 0.07 mV for Sweep Down
    	
    16:38  Field -> Vert +
    
    16:43  NMR Feq Sweep @ Vert
    	Data ID : NMR_Freq_He_20090111_164306
    	11.15 +- 0.06 mV for Sweep Up
    	-10.99 +- 0.11 mV for Sweep Down
    
    16:45  NMR Feq Sweep @ Vert
    	Data ID : NMR_Freq_He_20090111_164527
    	10.94 +- 0.07 mV for Sweep Up
    	-10.86 +- 0.09 mV for Sweep Down
    
    16:38  Field -> Trans -
    
    16:49  NMR Feq Sweep @ Trans
    	Data ID : NMR_Freq_He_20090111_164945
    	-17.79 +- 0.06 mV for Sweep Up
    	17.93 +- 0.07 mV for Sweep Down
    
    ===> Transverse Pickup Coil is ~20% noisier than Vertical one
    ===> Total Transverse NMR signal is 61% of that of Vertical one (It's 
    reasonable, considering: one of the 2 trans. pickup coil is working, 
    however the transverse coils are closer to cell)
    ===> Signal Noise Ratial of Transverse Coils is 50% of that of Vertical
    
    ===================================
          NMR Cross Calibration
    ===================================
    
    17:00  NMR Freq Sweep @ Trans
    	Data ID : NMR_Freq_He_20090111_170004
    
    17:01  Field -> Long. +; Target -> Pick up
    
    17:07  NMR Field Sweep @ Pickup
    	Data ID : NMR_He_20090111_170721
    
    17:08  Field -> Trans -; Target -> he3
    
    17:14  NMR Freq Sweep @ Trans
    	Data ID : NMR_Freq_He_20090111_171422
    
    
    ===================================
          Spin Flip via Field Rot.
    ===================================
    17:24  1st Field Sweep @ Trans -
    	Data ID : NMR_He_20090111_172450
    	average of both NMR signal is 11.8mV
    
    17:30  2nd Field Sweep @ Trans -
    	Data ID : NMR_He_20090111_173022
    	average of both NMR signal is 11.5mV
    
    17:31  20 Field Spin Flip, 45Degree Step.
    17:33  30 Field Spin Flip, 45Degree Step.
    
    17:36  3rd Field Sweep @ Trans -
    	Data ID : NMR_He_20090111_173022
    	average of both NMR signal is 11.4mV
    
    17:37  50 Field Spin Flip, 45Degree Step.
    
    17:42  4th Field Sweep @ Trans -
    	Data ID : NMR_He_20090111_173022
    	average of both NMR signal is 11.25mV
    
    17:44  1 Field Spin Flip, 45Degree Step.
    
    17:44  5th Field Sweep @ Trans +
            Confirm Spin direction is reversed
    	average of both NMR signal is -11.1mV
    
    17:45  1 Field Spin Flip, 45Degree Step.
    
    17:46  6th Field Sweep @ Trans -
    	Data ID : NMR_He_20090111_174652
    	average of both NMR signal is 10.7mV
    
    ===> Polarization Loss Rate of Spin Flip Using Field Rotation is < 
    200ppm
    ===> He3 Spin Follow field direction well during field direction jump 
    with same 25G field strenth (should follow up to 10kHz field change for 
    25G holding field)
    ===> Could be a useful spin flip technique for future experiment with 
    high rate spin flip and higher field gradiant. 
    
    ===================================
          Calib Field Sweep @ He3 Location
    ===================================
    17:56 Laser Back
    
    18:42  Freq Sweep before Field Sweep
    	Data ID : NMR_Freq_He_20090111_184251
    	-15.04mV sweep up
    	15.06mV sweep down
    
    18:43  Field Sweep
    	Data ID : NMR_He_20090111_184321
    	10.91mV Sweep up
    	10.78mV Sweep down
    	
    
    18:44  Freq Sweep after Field Sweep
    	Data ID : NMR_Freq_He_20090111_184251
    	-14.79mV Sweep up
    	14.73mV Sweep Down
    
    ===> Calibration Constant for Field Sweep is 3.83%/mV
    ===> Top Polarization of Last Spin up is 65.7%, spin up time = 3.82h
    
    
    
    
    