User name Jin
Log entry time 19:45:57 on January 11, 2009
Entry number 255483
This entry is a followup to: 255301
Followups:
keyword=Target Tests / Measurement
Couple measurement Has been performed on target this evening:
1. Trans. Pick Up Coil Performance Test:
Signal Noise Ratial of Transverse Coils is 50% +- 5% of that of
Vertical ones. (20% noisier and 40% less sensitive)
2. A NMR Cross Calibration is performed
3. Spin Flip Using Field Rotation is tested to be working great
Polarization Loss Rate of Spin Flip Using Field Rotation is < 200ppm
Currently take 1s to flip. Could finish flip on the level of
milliseconds.
Should not be sensitive to field gradiant
4. Top Polarization of Last Spin up is 65.7%, spin up time = 3.82h
Started at Halog 255301
http://www.jlab.org/~adaq/halog/html/0901_archive/090110161506.html
--------------------- Time Line & Details ---------------------
Start:
Cell Brady, Oven @ 230C
3x Comet Laser pumping and several test after that
5x amplication for all preamplifier
16:14 Laser OFF
===================================
Trans. Pick Up Coil Performance Test
===================================
16:37 NMR Freq Sweep
Data ID : NMR_Freq_He_20090111_163727
-18.02 +- 0.07 mV for Sweep Up
18.3 +- 0.07 mV for Sweep Down
16:38 Field -> Vert +
16:43 NMR Feq Sweep @ Vert
Data ID : NMR_Freq_He_20090111_164306
11.15 +- 0.06 mV for Sweep Up
-10.99 +- 0.11 mV for Sweep Down
16:45 NMR Feq Sweep @ Vert
Data ID : NMR_Freq_He_20090111_164527
10.94 +- 0.07 mV for Sweep Up
-10.86 +- 0.09 mV for Sweep Down
16:38 Field -> Trans -
16:49 NMR Feq Sweep @ Trans
Data ID : NMR_Freq_He_20090111_164945
-17.79 +- 0.06 mV for Sweep Up
17.93 +- 0.07 mV for Sweep Down
===> Transverse Pickup Coil is ~20% noisier than Vertical one
===> Total Transverse NMR signal is 61% of that of Vertical one (It's
reasonable, considering: one of the 2 trans. pickup coil is working,
however the transverse coils are closer to cell)
===> Signal Noise Ratial of Transverse Coils is 50% of that of Vertical
===================================
NMR Cross Calibration
===================================
17:00 NMR Freq Sweep @ Trans
Data ID : NMR_Freq_He_20090111_170004
17:01 Field -> Long. +; Target -> Pick up
17:07 NMR Field Sweep @ Pickup
Data ID : NMR_He_20090111_170721
17:08 Field -> Trans -; Target -> he3
17:14 NMR Freq Sweep @ Trans
Data ID : NMR_Freq_He_20090111_171422
===================================
Spin Flip via Field Rot.
===================================
17:24 1st Field Sweep @ Trans -
Data ID : NMR_He_20090111_172450
average of both NMR signal is 11.8mV
17:30 2nd Field Sweep @ Trans -
Data ID : NMR_He_20090111_173022
average of both NMR signal is 11.5mV
17:31 20 Field Spin Flip, 45Degree Step.
17:33 30 Field Spin Flip, 45Degree Step.
17:36 3rd Field Sweep @ Trans -
Data ID : NMR_He_20090111_173022
average of both NMR signal is 11.4mV
17:37 50 Field Spin Flip, 45Degree Step.
17:42 4th Field Sweep @ Trans -
Data ID : NMR_He_20090111_173022
average of both NMR signal is 11.25mV
17:44 1 Field Spin Flip, 45Degree Step.
17:44 5th Field Sweep @ Trans +
Confirm Spin direction is reversed
average of both NMR signal is -11.1mV
17:45 1 Field Spin Flip, 45Degree Step.
17:46 6th Field Sweep @ Trans -
Data ID : NMR_He_20090111_174652
average of both NMR signal is 10.7mV
===> Polarization Loss Rate of Spin Flip Using Field Rotation is <
200ppm
===> He3 Spin Follow field direction well during field direction jump
with same 25G field strenth (should follow up to 10kHz field change for
25G holding field)
===> Could be a useful spin flip technique for future experiment with
high rate spin flip and higher field gradiant.
===================================
Calib Field Sweep @ He3 Location
===================================
17:56 Laser Back
18:42 Freq Sweep before Field Sweep
Data ID : NMR_Freq_He_20090111_184251
-15.04mV sweep up
15.06mV sweep down
18:43 Field Sweep
Data ID : NMR_He_20090111_184321
10.91mV Sweep up
10.78mV Sweep down
18:44 Freq Sweep after Field Sweep
Data ID : NMR_Freq_He_20090111_184251
-14.79mV Sweep up
14.73mV Sweep Down
===> Calibration Constant for Field Sweep is 3.83%/mV
===> Top Polarization of Last Spin up is 65.7%, spin up time = 3.82h
A copy of this log entry has been emailed to: yqiang, jpchen, chiranjib