    Make New Entry, Make Followup Entry

    User name brads

    Log entry time 12:11:23 on January 13, 2009

    Entry number 255837

    This entry is a followup to: 255461

    keyword=BB Cerenkov runlist for gain check of Cerenkov PMTs 1--10 (late entry)

    Summary: Juggling cables at the FE amplifier seems to have 'solved' the intermittent noise problem on PMTs 2,3. PMTs 1--10 (beamline side) gain matched to put peak at ADC channel 30.

    Trigger T2 was restored to be the Cer+Sh overlap trigger. ADC gate timing with this trigger was double checked at the Cerenkov PMTs. Looks good.


    Run List
    Run 4910:
      - HVs for 1--10 on, pulser on
      - LED #16, #20 plugged in
        - Pulser amplitude sent to LED is 1.25V, 25ns wide
      - Pedestal run
        - T8 trigger dominated (others prescaled away)
      = Pedestals mostly OK
        - still noise on ADC chans 1,2
        - similar noise on ADC chans 17,18.  Those PMTs are powered off right now
          - this noise was NOT there yesterday...
    
    Run 4911:
      - identical config to 4910
      = same results as 4910
    
    
    Run 4912:
      - HVs for 1--10 on, pulser on
      - cables to LED #16, #20 have been unplugged
        - No LEDs are being powered right now.
      = same noise as runs 4910--11
    
    [ Lots of F1 FIFO overflow errors on replay this afternoon.
      - tried power cycling ROC5 crate -- didn't help ]
    
    Run 4913:
      - same config as 4912, but ROC5 was rebooted
      - HVs for 1--10 on, pulser on
      = same results
    
    Run 4914:
      - same config as 4912
      - tried reseating the cables for those PMTs at the amplifier (since we had
        looked at the signal from PMT 16,17 earlier today)
      = same results
    
    
    
    Run 4915:
      - HVs for 1--10 on, pulser on
      - pedestal run using LED pulser trigger
        - NOTE: cables to LED #16, #20 are still unplugged
        - No LEDs are being powered right now.
      - changed prescales so the DAQ is triggered on the pulser-driven T2 trigger now
      = same results as with T8 trigger (good)
        - noise still there though...
    
    
    Run 4916:
      - HVs for PMTs 1--10 on, pulser on
      - cables to LED #16, #20 plugged back in.  LEDs are firing now
        - Pulser amplitude sent to LED is 1.25V, 25ns wide
      = after run we adjusted HVs on PMTs 5,7--10
        - PMTs 1,4,6 look good
        - PMTs 2,3 (noisy) are hard to tell
    
    Run 4917:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is now 1.35V, 25ns wide
      = after run we adjusted HVs on PMTs 5,7--10
        - PMTs 1,4,6 look good
        - PMTs 2,3 (noisy) are hard to tell
    
    Run 4918:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      = after run we adjusted HVs on PMTs 5,9--10
        - PMTs 1,4,6,8 look good
        - PMTs 2,3 (noisy) are hard to tell
    
    Run 4919:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      = after run we adjusted HVs on PMT 4
        - PMTs 1,5--10 look good
        - PMTs 2,3 (noisy) are hard to tell
    
    Run 4920:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      = PMTs 1,4--10 look good
        - PMTs 2,3 (noisy) are hard to tell
    
    Run 4921:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      - Moved the signal cable for PMT 2 from its amplifier channel and plugged it
        into that for PMT 11 (different amplifier, different crate)
      = Noise on ADC ch 1,2 is now gone.
        - ADC ch 1 (no signal now) has sharp, good pedestal
        - ADC ch 2  (PMT 3) looks quite good
        - ADC ch 10 (PMT 2) looks quite good
    
    Run 4922:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      - signal cable for PMT 2 from its amplifier channel is still in that for PMT
        11 (different amplifier, different crate)
      - unplugged top ribbon cable from F1 TDC in ROC5, slot 11 to see if the F1
        overflow errors went away
      = Overflow errors did NOT go away
       
    Run 4923:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      - signal cables for PMT 2 are now plugged into an unused channel in its original
        amplifier
      - PMT 11 is restored
      - plugged top ribbon cable back in, removed _bottom_ ribbon cable from F1 TDC
        in ROC5, slot 11 to see if the F1 overflow errors went away
      = Overflow errors DID go away
        - PMTs 2,3 both look good now.
        - noise on ADC channels for PMTs 18,19 is gone now too...
       
    Run 4924:
      - HVs for 1--10 on, pulser on
      - Pedestal run again
        - T8 trigger dominated (others prescaled away)
      = all noise still gone, all peds look good
    
    Run 4925:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      - (Now gain check PMTs 2,3)
      = Raised HV on PMT 3
        - PMT 2 is good
    
    Run 4926:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      - (Now gain check PMT 3)
      = lowered HV on PMT 3
    
    Run 4927:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      - (Now gain check PMT 3)
      = lowered HV on PMT 3
    
    Run 4928:
      - HVs for PMTs 1--10 on, pulser on
      - Pulser amplitude sent to LED is 1.35V, 25ns wide
      - (Now gain check PMT 3)
      = PMT 3 is now OK