• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name A. Camsonne

    Log entry time 10:57:43 on January30,2009

    Entry number 258314

    keyword=Compton Electron detector control board moved

    Compton Electron detector control board moved to other slow control crate.
    I was fearing that this board in A16 could interfere with other boards on the crate. It seems this was confirmed since removing this board allowed to lock the cavity. So the board was moved back to the other IOC, I guess looking at the adresses available could fix it or I will see if we can get an A24 board for the controller of the motion.