    User name Kalyan

    Log entry time 12:49:17 on February 02, 2009

    Entry number 258692

    This entry is a followup to: 258689

    Followups:

    keyword=Re: BB gas not bubbling?

    There are four bubblers on the panel, two for the input line(higher level of liquid) and two for the output lines. The output lines should bubble all the times but input lines will not. If non of them are bubbling then there might be a problem. Need to check in the gas shed to see if flow is still good. The gas is ethane-argon mix.