Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name jpchen/yiqiang
Log entry time 16:12:36 on February 06, 2009
Entry number 259271
keyword=pressure gauge/cooling jet flow
We calibrated the pressure gauge and vacuum gauge for the reference cell
system. Zeroed the vacuum and set pressure gauge to -15 psig (both CH and
hall) when the local vacuum gauge reading 7 mtorr. Then put in 1 atm N2
and checked that the pressure gauge read 0 psig and vacuum read 773/774 torr.
Requested Cryo group to turn on He supply and check our cooling jet for
the target cell windows are set at 40 and the flow to the beam Be window
at 2.5.