Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name O. Glamazdin
Log entry time 21:45:47 on February 11, 2009
Entry number 260554
keyword=Moller results
Moller results on Feb 11 15:25-21:10
Mini Spin Dance
- There was no energy lock..
- The raster was ON.
- For the beam energy current - see the chart attached.
Preliminary results: not corrected for the beam
energy fluctuations.
# |
Wien angle |
beam |
runs |
target |
Ptarget% |
Pbeam% |
1 | 30° | 0.8 µA
| 13174-13175 | 5 | 8.03 |
81.5±0.2±2.0 |
2 | 30° | 0.8 µA
| 13176-13179 | 4 | 7.51 |
81.1±0.2±2.0 |
Aver. | 30° | 0.8
µA | 13174-13179 | 4,5 | - |
81.2±0.1±2.0 |
1 | 20° | 0.8 µA
| 13180-13181 | 5 | 8.03 |
72.6±0.2±2.0 |
2 | 20° | 0.8 µA
| 13182-13185 | 4 | 7.51 |
70.5±0.2±2.0 |
Aver. | 20° | 0.8
µA | 13180-13185 | 4,5 | - |
71.1±0.1±2.0 |
1 | 23° | 0.8 µA
| 13186-13187 | 5 | 8.03 |
73.2±0.2±2.0 |
2 | 23° | 0.8 µA
| 13189-13190 | 4 | 7.51 |
73.5±0.2±2.0 |
Aver. | 23° | 0.8
µA | 13186-13190 | 4,5 | - |
73.3±0.1±2.0 |
Spin Dance Results
The Wien filter angle 23° was choosen as an optimal for the Hall
A and the Hall C beam polarization values.
Hall A Moller results are presented in the chart attached.
All results are preliminary only and have to be corrected on the beam
energy fluctuation.
A copy of this log entry has been emailed to: gen@jlab.org
Figure 1

Figure 2