keyword=reference system gauges re-calibrated.
Recalibrated the pressure and vacuum gauges for the reference cell system:
Before calibration, at 1 atm (after venting), Upstairs pressure reading was -13 psig, vaccum gauge reading was 769 torr; at vacuum,
pressure reading was -28 psig and vaccum reading was -3 torr; the downstair pressure gauge reading also -28 psgi and vaccum reading -5 torr.
Zeroed both vaccum gauges and changed both pressure gauge to be -15 psig. Put to 1 atm N2, Downstair vaccum gauge reading was 772 torr
and upstair reading 773 torr. Both pressure gauge readings were 0 psig.