Log entry time 23:38:35 on March 01, 2009
Entry number 264148
This entry is a followup to: 264138
Followups:
keyword=JLab laser over temperature; check output; power on again
JLab COMET laser had over-temperature tripped when I came to counting
house. It was checked and powered on again. Full power tested for half
hour, it seems working normally now.
There is some dust on JLab laser jumper cable tip (the end connecting to
long fiber). It's cleaned.
Then, the JLab laser output power curve is tested to be normal:
10A -> 1.18W
20A -> 8.99W
30A -> 17.1W
36A -> 22.5W <= Working condition
40A -> 25.6W
Then the trip temperature threshold of JLab laser was increased by 1
degree. Current trip threshold setting and diode temperature read back
(Taux in diode controller) are:
Laser Temp. Threshold
JLab 27.2C 31C (increased from 30C)
W&M 29.0C 30C
Rugster 24.0C 27C
Jumper fiber of JLab laser was replaced (Halog 262886) one week ago,
which could caused diode working temperature higher than before.
A copy of this log entry has been emailed to: yqiang, jpchen, riordan