    User name Yi Qiang

    Log entry time 21:24:47 on March 05, 2009

    Entry number 264784

    This entry is a followup to: 264782

    Followups:

    keyword=Readings for Reference cell

    The reference cell temperatures are recorded to the data stream all the time, however the gas type and pressure are not. As for the pressure, because the temperature of the cell is depend on the location and beam condition, the in beam pressure of the cell will be slightly different (1~2%) from the pressure recorded in other conditions. Among all these readings, the in-beam pressure is the most reliable one.

    Also, keeping a manual log of gas type will give the one who sort out the run list at the end a chance to cross check with the automatic run information and rates.

    As a Target operator, if he/she is not sure about the gas type in the reference cell, by getting permission from RC, it's strongly recomended to vent/vacuum the cell and fill the cell with proper gas.

    Sorry that I didn't put these stuff into the TO check list, and I will update the check list with these new information.