• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name Moffit

    Log entry time 18:27:30 on May 01, 2009

    Entry number 270288

    This entry is a followup to: 270285

    keyword=BCM dx3: removed patch cable... problem goes away.

    I removed the patch cable that we put in, as described in the previous HALOG. And the craziness went away.

    Attached is the happex7 scaler reading of the BCM dx3 and ux3 combination. If everything is good, then the difference should be zero... and the quotient should be 1. After removing the cable, it looks "good" (by the previous definition).

    This patch provided a fanout of BCM Dx3 to a cable labeled as going to Compton. I wonder if the end of this cable is actually plugged in (goes to another patch panel that goes down to the Hall).




    Figure 1