Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name V. Sulkosky
Log entry time 18:41:45 on May 03, 2009
Entry number 270675
This entry is a followup to: 270660
keyword=Left HRS Gas Cerenkov #6
This behavior of this channel changing amplitude was first noticed before
the end of Transversity. I attempted to fix it by switching delay
cables, which did not work. I then switched the HV channel on the 1461N
board for the gas Cerenkov. For d2n, we changed kinematics quite a bit,
but sometimes we ran at the same kinematic for a few days. I was never
informed of any issues with this channel during that experiment.
It seems the changing amplitude has come back or maybe was never gone in
the first place.
It symptom appears to be a bad cable or connection somewhere. I do not
believe that the PMT itself has a problem.
Unfortunately when I attemnpted to check the problem before, the
amplitude was normal. Someone would need some dedicated time during
beam studies to track down the issue with this channel.
A copy of this log entry has been emailed to: yawei,zhaobo,peter