Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name Yi Qiang
Log entry time 12:19:00 on May 05, 2009
Entry number 270996
This entry is a followup to: 269713
keyword=EPR Calibration for Vertical Pumping
== Test Condition ==
Cell: Dominic
Oven Temperature: 230C
Field Direction: Transverse -
Laser Power: 3x25 COMET
Cell Density: 10.83 amg
== Rb Parallel ==
Pumping Direction: Parallel
Data directory: C:\AyeD\EPR\Data\EPR_AFP_20090505_1159
Calibration Constant: 3.27 %/mV
Polarization: 63.5%
Field: 25.04 G
EPR Frequency Shift: 85.53 kHz
== K Parallel ==
Pumping Direction: Parallel
Data directory: C:\AyeD\EPR\Data\EPR_AFP_20090505_1146
Calibration Constant: 2.89 %/mV
Polarization: 57.5%
Field: 25.08 G
EPR Frequency Shift: 90.31 kHz
== Rb Anti-Parallel ==
Pumping Direction: Parallel
Data directory: C:\AyeD\EPR\Data\EPR_AFP_20090505_1117
Calibration Constant: 2.76 %/mV
Polarization: 57.0%
Field: 25.07 G
EPR Frequency Shift: 82.98 kHz
== K Anti-Parallel ==
Pumping Direction: Anti-Parallel
Data directory: C:\AyeD\EPR\Data\EPR_AFP_20090505_1125
Calibration Constant: 2.92 %/mV
Polarization: 59.0%
Field: 25.06 G
EPR Frequency Shift: 146.03 kHz
The Calibration constants from Rb is still not consistent with K and they
don't even agree with each other in different pumping direction. So I
will use the most accurate calibration from K parallel measurement as the
new constant.
A copy of this log entry has been emailed to: jpchen, tdaver@wm.edu, miham, yawei, jinhuang
Rb Anti-parallel AFP

Rb Anti-parallel Analysis

K Anti-parallel AFP

K Anti-parallel Analysis

Rb Parallel AFP

Rb Parallel Analysis

K Parallel AFP

K Parallel Analysis