    User name miham

    Log entry time 13:49:29 on May 05, 2009

    Entry number 271003

    This entry is a followup to: 270370

    keyword=BB dE dead channel #9 fixed

    Today during the controlled access I was trying to fix the dead TDC
    channel #9 (#8 if we count from 0) in the dE-scintillation plane of
    the bigbite.

    First I checked if the signal from the amplifier comes to the input
    into the discriminator. I realized that there was no signal.
    Therefore I started tracing it back. When I was moving the output cables
    from the signal splitters (that split signals from the amplifiers into
    two copies one for the TDCs and one for the Trigger) the signal
    reappeared. It seems that the connector for this channel was not plugged
    well into the module. I have disconnected this channel, checked the
    connectors and plugged the cables back in. Afterwards I have taken a test
    run with cosmics and it seems, that the channel works again.


    A copy of this log entry has been emailed to: moffit, sirca, peter, miham



    TDCs



    ADCs