Log entry time 15:35:30 on May 05, 2009
Entry number 271009
This entry is a followup to: 270997
keyword=Detector Hut air conditioners
Right Detector Hut air conditioning unit was found to be powered off. It
is now on and cooling normally.
LEFT Detector Hut air conditioner was dripping water from an over filled
drip tray. The MRI crew discovered the drain line was elevated at the
elbow near the back of the hut. A new elbow was installed and the drain
line slope was corrected. The leaking unit is now fixed.
Due to the quick response of the MRI crew the problem was fixed by 3:00
P.M. THANKS to all!
Shield House doors are closed. Master key returned.