NEXT
Make New Entry,
Make Followup Entry
User name R. Michaels / D. Wang
Log entry time 22:20:33 on May 5,2009
Entry number 271158
keyword=Compton test runs
Start 21:30
Runs with the new Compton DAQ configs and normal beam conditions.
Run 16814 .... Prod_FA config .... no beam at start of run
Run 16815 .... Prod_FA
Switch to Prod_FT
Run 16816 .... Prod_FT config
Finish 22:15 ... Return to normal Production, start run 16817.
I'd like the experts to check these runs.
For convenience I copied them to /work/halla/parity/disk1/bob
Observations :
1. The Compton logbook shows normal results for run 16814, so
the FADC data probably does no harm. Run 16815 may have been
too short. 16816 looks ok too.
2. Running the Compton analysis by hand, we see lots of complaints
about "unknown ROC6" -- trivial to fix. ROC6 is new FADC crate.
3. CMU analysis code needs a simple change in mainanal.cc:
pROCdata=datafile.getROCdata(6,MyDebug);