    User name V. Sulkosky

    Log entry time 15:33:30 on May 06, 2009

    Entry number 271258

    This entry is a followup to: 270981

    keyword=Re: L-HRS Gas Cerenkov ADC # 6

    I decided to move the left HRS Gas Cerenkov ADC # 6 channel input to the
    P\S 776 amplifier from channel 7 to channel 11. This will test to see
    if the problem is due to the amplifier. I put a pulser signal into
    channel 7 and found that the amplification was x10, which is what I expect
    it to be. On the other hand, this problem is intermittent, so I may be
    unfortunate in my timing.

    The first run with this change is 1894. This doesn't affect the mapping
    in any way.

    A copy of this log entry has been emailed to: zhaobo,yawei,yqiang