    User name Higinbotham

    Log entry time 16:18:30 on May 8,2009

    Entry number 271755

    keyword=BCM Calibration Check After Change


    After the change of the 5V supply, beam current is stable but was reporting a bit high.

    Old calibration constants: 48.345 & 47.773
    New calbiration constants: 45.45 & 44.6

    I will check again after the 4:15 meeting, but this should be good for the weekend. Hopefully it will just be stable as typically we NEVER change these constants.