    User name V. Sulkosky

    Log entry time 15:18:09 on May 11, 2009

    Entry number 272599

    This entry is a followup to: 272555

    Followups:

    keyword=Re: BB triggers to RHRS on long ribbon cable (grey band)

    I connected the ribbon cable to a LeCroy 4616 NIM-ECL converter.
    However, this module requires that the red wire be on the top (1st pair), so I
    flipped the cable at the BigBite weldment from the output of the P\S 707 discriminator.

    After wards, I found that channels 1 and 3 did not show a signal. There
    is a signal from channels 2 and 4, but the width is rather narrow. I will
    need to use another module to widen the signals so that the scalers will count the pulses.
    Here is the scaler map for these triggers on the R-HRS:
    Trigger     Norm Channel     Chan Name
    +++++++++++++++++++++++++++++++++++++++
    T1               26            T1_BB
    T2               27            T2_BB
    T5               30            T5_BB
    T6               31            T6_BB
    


    A copy of this log entry has been emailed to: zhaobo,moffit,miham