Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name R. Michaels
Log entry time 17:55:16 on May11,2009
Entry number 272622
keyword=scaler moved to Compton FADC crate
The CAEN scaler that was at the far right of the formerly used
MWPC DAQ in the lower VME crate was moved to the Compton FADC crate,
with Brad's permission.
The scaler appears in the datastream with header dae56000 and is
at the end of the regular scaler bank. It looks like the analyzer
does not get confused by this.
The 40 MHz clock is in the 4th channel and tracks the other scaler.
I did not touch the other scaler. It has been working all after-
noon since the power-cycle / re-seating.
Now there is a new disease: The FADC crate constantly reports
interrupt: FADC not busy state= 1
I suppose something got unplugged. Will check tomorrow.