    User name Jack (Techs)

    Log entry time 19:21:33 on May 16, 2009

    Entry number 273949

    keyword=Right Q1

    It would appear we have lost communication with the Right Q1 power supply. What I would like to do is get a chance to go into the hall and look at the supply directly with a laptop, to see if the supply is hung or the EPICS serial interface is hung. Whichever, it would be a serious setback if someone just went into the hall and reset the power supply without checking this out. I have attached a jpg of the RQ1 GUI taken at 19:11 today. Since the field seems to be stable I would say we could do an access when convenient, but I would give it some priority.