    User name Moffit / miham

    Log entry time 15:55:53 on May 19, 2009

    Entry number 274812

    keyword=T2 double pulsing problem solved

    Today during the access Bryan and myself went down the hall
    to solve the problem with T2 double pulsing in the HRS-L
    scaler. We tried various things in order to solve the problem.
    First we were swapping the channels (cables). At one moment
    problem seemed to disappear. However after few minutes the
    double pulsing reappeared. In the next step we unplugged
    the T2 and T6 scaler signals from the trigger electronics and
    plugged the 1kHz pulsar directly into them. When we checked the
    the scaler rates in HRS-L and Bigbite GUI, they matched. We were
    clueless. Since everything seemed to be working again, we plugged
    all the channels back to the original configuration, but once we did
    that double pulsing again came back. After that we have decided to
    change module LeCroy 4616 with equivalent Phillips module. First
    it seemed that this helped but when we plugged-in all the channels,
    the problem came back. At that point Bryan realized that when he unplugs
    the T1, the problem disappears. He suspected on the cross-talk between
    the T1 and T2.

    After we realized what the problem was, we went up into the HRS-L and
    played with the cables a bit. We localized the problem to the Daisy chain
    ribbon cable leading to the scalers. When all the connectors were
    plugged into the modules the T2 was double pulsing. If Bryan unplugged
    one of them (it did not matter which one) the problem disappeared. At
    that point we have decided to change this cable. We made a new ribbon
    cable, tested it, and plugged it into the scalers. After that, problem
    seemed to go away. Now (at least with the single frequency pulser)
    all scalers look fine.