Log entry time 12:44:51 on May 21, 2009
Entry number 275294
keyword=AFP Loss test for Transverse+ field
The AFP Loss was tested with different correction coil current settings.
The polarization is relatively low <40% so masing effect can be ignored.
Data files Small Large AFP Loss
C:\AyeD\NMR\Data\NMR_Freq_20090521_115542 +1.0A +0.0A 1.25%
C:\AyeD\NMR\Data\NMR_Freq_20090521_121822 +0.5A +0.0A 1.04%
C:\AyeD\NMR\Data\NMR_Freq_20090521_121036 +0.0A +0.0A 0.82%
C:\AyeD\NMR\Data\NMR_Freq_20090521_120632 -0.5A +0.0A 0.94%
C:\AyeD\NMR\Data\NMR_Freq_20090521_120238 -1.0A +0.0A 1.25%
C:\AyeD\NMR\Data\NMR_Freq_20090521_122302 +0.0A +2.0A 1.10%
C:\AyeD\NMR\Data\NMR_Freq_20090521_122838 +0.0A -2.0A 1.08%
So clearly, the field gradient is minimal without any correction field.
So I will leave the small correction ON with 0.5A which gives roughly
15mG/cm gradient to see whether this will help us to increase the
polarization.