Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name Yi/Jin
Log entry time 22:05:22 on May 22, 2009
Entry number 275598
Followups:
keyword=Target activities in the hall
We did the following check in the hall:
1) Transverse laser alignment: we adjusted the laser spot slightly (~0.5cm)
2) Transverse rotatable Q-wave plates angle: moter#1474 is very good,
motor#0338 got re-calibrated and the change is 1.7 degrees.
3) Laser power measurement: 57.6 W in total (no big change)
Even the additional gradient from the small correction coil didn't help,
we decided to leave it on (0.5A) to reduce the masing effect for future
EPR measurement.
A copy of this log entry has been emailed to: jpchen, doug, miham