    User name Jin Huang

    Log entry time 22:26:48 on May 22, 2009

    Entry number 275600

    keyword=1/4 plates calibration for Transverse Laser

    Motorized quarter wave plates #3 and #4 (Transverse S wave and P wave) have been recalibrate in target lab. The result shows the physical optimum angle is consistent within 1 degree from previous calibration in Summer 2008, which means the plate has been attached well with its holder.

    However, the edge angle of limit switch of motor #3 has shifted by 1.7 degree (due to mechanical worn-out?). Software is using this edge to synchronize software angles and physical 1/4 plates angles. Therefore, transverse line S wave plate has been 1.7 degree off from supposed optimum ("supposed optimum" could also be ~1 degree off from physical optimum angle due to calibration limitation).

    But, according to previous operation, we know that 1~2 degree offset on 1/4 plate will not significantly limit max target polarization.

    The new limit switch position is saved to 1/4 plate control code. New setting are:

    motor #0338 = plate #3 : Up Pos = -82.7 Degree; Down Pos = 7.3 Degree

    Other motors are not changed