User name ellie
Log entry time 15:35:30 on May 28, 2009
Entry number 277119
keyword=HAND Access Activities, HV Changed
I went into the Hall with Doug to see if we could figure out what was
wrong with V-22L, N3-11L, N3-12L, N3-11R, and N3-12R.
V-22L :
· Looked at signal on o-scope
→ Looked good if a little slower rate
compared to the other signals
· Unplugged signal cable from Amplifier and put V-23L in
its place
→ V-23L looked good
→ Put V-23L back where it belongs
· Checked V-22L's amplifier channel with a pulser
→ Pulsed signal coming out of
amplifier looks good
· Determined that the amplifier was not the problem
→ Put V-22L signal back into amplifier
· Increased V-22L HV to 1250 V
→ This increased the rates similar to
other bars, we'll see if it helps
Plane 3 :
· All of the problematic bars in Plane 3 were after pulsing
· The voltages of each were lowered by 150 V so that the
after pulses would be below the threshold (-50mV)
→ N3-11L is now at -1600 V (was -1750
V)
→ N3-12L is now at -1555 V (was -1705
V)
→ N3-11R is now at -1665 V (was -1850
V)
→ N3-12R is now at -1600 V (was -1750
V)