    User name ellie

    Log entry time 15:35:30 on May 28, 2009

    Entry number 277119

    keyword=HAND Access Activities, HV Changed

    I went into the Hall with Doug to see if we could figure out what was wrong with V-22L, N3-11L, N3-12L, N3-11R, and N3-12R.

    V-22L :
    · Looked at signal on o-scope
         → Looked good if a little slower rate compared to the other signals
    · Unplugged signal cable from Amplifier and put V-23L in its place
         → V-23L looked good
         → Put V-23L back where it belongs
    · Checked V-22L's amplifier channel with a pulser
         → Pulsed signal coming out of amplifier looks good
    · Determined that the amplifier was not the problem
         → Put V-22L signal back into amplifier
    · Increased V-22L HV to 1250 V
         → This increased the rates similar to other bars, we'll see if it helps


    Plane 3 :
    · All of the problematic bars in Plane 3 were after pulsing
    · The voltages of each were lowered by 150 V so that the after pulses would be below the threshold (-50mV)
         → N3-11L is now at -1600 V (was -1750 V)
         → N3-12L is now at -1555 V (was -1705 V)
         → N3-11R is now at -1665 V (was -1850 V)
         → N3-12R is now at -1600 V (was -1750 V)