    keyword=Target Ladder Stuck

    When the spin rotation finished, we tried to move the target to ref cell position. The ladder motion GUI show the brake is OFF. However, the limit switch on the "He3 position" do not go off and the laser spectrum keep unchanged.

    Target IOC 26 is rebooted for 3 times, the problem persists. We decide to have a control access to reboot the motor controller and check the connectors.

    It looks like the brake is not off, like what has happened in the beginning of Transversity

    A copy of this log entry has been emailed to: yqiang, jpchen, doug, camsonne