    Log entry time 13:07:54 on June 06, 2009

    Entry number 279394

    This entry is a followup to: 277446

    Followups:

    keyword=Target spin rotated to Transverse - (270 degrees)

    Target Spin was rotated to Transverse - (270 degrees), while the field was rotated to Transverse + (90 degree).

    1. Stop Laser
    

    2. Rotate Field to Transverse +

    3. Setup 1/4 plates 0338/1474 angles following to Spin UP (Transverse -)

    4. Swap fibers to Transverse Pumping: Jlab->N2, W&M->N4, Rutgers->N5

    7. Power on laser

    8. Swap Pickup Coil Cable(4->1); Change Calibration Constant (1.949x2.5); Change NMR Default settings

    9. NMR @ Transverse-, found out the calibration constant should not be multiplied by 2.5 since the pre-amplifier for the longitudinal pickup coils has never been changed, so change the Calibration Constant back to 1.94902

    10. NMR @ Transverse - with correct calibration constant: 56.5%






    Field Setting



    Motor 0338



    Motor 1474



    NMR with incorrect Calibration Constant



    NMR