Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name Yi/Miha
Log entry time 18:51:31 on June 10, 2009
Entry number 280444
This entry is a followup to: 272648
Followups:
keyword=Replaced damaged jumper fiber for the comet lasers
During the time I was checking the fiber temperature after we switched to
the longitudinal pumping, I found out the connections between the jumper
fibers from Rutgers and Jlab lasers to the 75 long fibers were
very hot. After talked to RC, we decided to take a look by turning off
these lasers.
It turned out to be that the jumper of the Rutgers and Jlab laser were
badly damaged. We decided to change black jumper fiber cables (that lead
from the laser diode to the metal cables that are going to the Hall) for
these two lasers.
When we were attaching cables (from laser controller
and temperature controller) back to diode enclosure one of the screws on
the connector (on the cable from laser controller) broke. We decided to
replace this cable with the cable from the new, recently bought laser.
After that we also did the laser power test, to make sure, that lasers
are working properly. In the end we connected new fibers to the fibers
leading to the Hall and turned on the lasers.