• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name miham

    Log entry time 08:49:46 on June 14, 2009

    Entry number 281274

    keyword=Calibration runs #3584-#3587: Short summary

    Here is a short summary regarding the calibration runs with reference 
    cell, that we took in the end of the OWL shift:
    
      Gas  |  Required num of T3 ev. | T3 rate | Ps. T3 rate | Estimated time
    --------------------------------------------------------------------------
      3He  |            1M           |  7.5kHz |  1875Hz     |  9min      
     Empty |            0.5M         |  4.5kHz |  1125Hz     |  7.4 min
      H2   |            2M           |  7.3kHz |  1825Hz     |  18min
      N2   |            2M           |  9.2kHz |  2300Hz     |  15min
    
    Prescale factor for T3 was set to PS3 = 4. The beam current was 
    approx. 14uA. 
    
    In order to get the correct number of T3 events for each runs we
    considered beam trips and appropriately extended the length of each
    run. Therefore are runs few minutes longer than planned.     
    
    #3584 : 3He  : 12min
    #3585 : Empty: 16min
    #3586 : H2   : 23min
    #3587 : N2   : 20min   
    
    


    A copy of this log entry has been emailed to: doug