    keyword=PREX Quartz off or unplugged

    PREX Quartz does not appear in ADCs or TDCs.

    Its supposed to be in the 2nd Cerenkov channel (R-HRS).
    The 1st and 3rd channels (HAPPEX and PREX-scint) do
    show up, so there isn't anything grossly wrong at the
    discriminator.

    The xscaler display is showing ~0.3 Hz at 20 uA but at that level
    it could simply be crosstalk from an adjacent channel.

    Tentative conclusion is that either the HV is off or the
    signal is not plugged into the discriminator. I'd check
    there on a scope first.