Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name O.Glamazdin
Log entry time 23:02:59 on August 31, 2009
Entry number 287270
Followups:
keyword=The Moller spin dance results
Moller results on August 31 14:50-23:50
Spin Dance
# |
Wien angle |
beam |
runs |
target |
Ptarget% |
Pbeam% |
1 | 44.30° | 0.55
µA | 13174-13175 | 5 | 8.03 |
-80.28±0.3±2.0 |
2 | 110.27° | 0.55
µA | 13174-13175 | 5 | 8.03 |
14.11±0.3±2.0 |
3 | -70.22° | 0.55
µA | 13174-13175 | 5 | 8.03 |
-5.37±0.3±2.0 |
4 | 14.20° | 0.55
µA | 13174-13175 | 5 | 8.03 |
-89.61±0.3±2.0 |
5 | 14.20° | 0.55
µA | 13174-13175 | 4 | 7.05 |
-87.85±0.3±2.0 |
6 | 14.20° | 0.36
µA | 13180-13181 | 3 | 7.56 |
-88.38±0.2±2.0 |
7 | 14.20° | 0.25
µA | 13182-13185 | 2 | 8.20 |
-89.06±0.2±2.0 |
Aver. | 14.20° | - |
13180-13185 | 2,3,4,5 | - |
-88.72±0.1±2.0 |
Spin Dance Results
The Wien filter angle 14.2° was choosen as an optimal for the
Hall A with the beam energy Eb=3.4839GeV (from M. Tiefenback) and linacs
disbalance=7.5%.
The Wien filter angle 17.° is an optimal for the Hall A with the
beam energy Eb=3.4812GeV and linacs disbalance=7.5%.
The beam polarization difference between 14.2° and 17° is
0.998682.
The Hall A Moller results are presented in the chart attached.
A copy of this log entry has been emailed to: gen@jlab.org, saha@jlab.org
Figure 1. Spin dance result

Figure 2. Beam polarization at 14.2deg Wien filter
