Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name R. Michaels
Log entry time 15:12:32 on September 04, 2009
Entry number 287988
This entry is a followup to: 287980
keyword=Rate problem understood
The clock on the R-HRS, which is in 8th channel of the norm scaler,
is quadrupole pulsing: 4*1024 Hz. This means the rates are actually
4 times higher.
The L-HRS clock looks ok.
I verified this at the level of raw data.
This brings the 2 HRS in line with each other and the L-HRS is
at 3 kHz for ~0.5 uA which is what we had before.