• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name paschke

    Log entry time 20:40:59 on September 5,2009

    Entry number 288174

    This entry is a followup to: 288171

    Followups:

    keyword=Surprising! results of dithering system test

    
    Below shows a typical beam modulation cycle.  in order, the beamline components 
    dithered are listed, along with color in the plot and the amplitude of the 
    triangle wave modulation.  This cycle is from run #12035.
    
    1) 1C08 H  - red - 100 mA current
    2) 1C08 V  - blue - 250 mA current
    3) 1C10 H  - red  - 250 mA current 
    4) 1C10 V  - blue - 100 mA current
    5) SL Zone 20  - green - 50 keV / pass set
    
    The plots show bpms vs. 30Hz sample number, starting with the top right, of 
    bpmx 12x, 4ax, 4ay, 4bx, 4by.
    
    Some conclusions:
    1) The energy setpoint of 50 keV /pass should correspond to 150 keV / 3484 MeV 
    ~ 4.3e-5.  But the 0.5mm modulation amplitude at bpm 12x (which has, I believe, 
    a dispersion of 4 meters) suggests that the energy modulation is 500e-6/4.0 = 
    1.25e-5.  This suggests that the energy kick is three times larger than the 
    setpoint would suggest.  Something to keep in mind.
    
    2) There is an apparent vertical dispersion at the target which is very large.  
    The modulation amplitude in response to energy was about 150 um, which (in 
    comparison the 500 um at 4meter dispersion in 12x) implies that the vertical 
    dispersion at the target is about 1.2 meters!
    
    3) The horizontal coils completely fail to span the beam phase space.  They 
    look as redundant as can be, which is pretty weird.  In fact, the amplitude is 
    similar in each bpm, which suggests that both coils produce position (but not 
    angle) deviations.  I can't figure this out: the 1C08 and 1C10 coils are 10 
    meters apart, with no active optics between them (or so I was told... I haven't 
    verified).  Thinking of an analogy to an optical system: it is as if each point 
    is at the focus, and winds up being well collimated by the following lenses, 
    even though the two points are 10 meters apart.  I'm no expert, but I find this 
    very bizarre.
    
    4) I might believe that the vertical is a little more independent, but it isn't 
    very convincing. 
    
    So, I'm surprised by the setpoint calibration (I remembered this setpoint 
    actually being close to the observed energy kick, in the past), the vertical 
    dispersion, and the lack of orthogonality in dithering magnets.  I believe that 
    the vertical dispersion is a sign of a potentially important deviation from 
    optical lattice design (and is notable given recent results suggesting a 
    mis-steer of the scattered photon beam from the Compton).  
    
    We will definitely require help from opticians to achieve orthogonality of the 
    horizontal and vertical correctors. 
    


    FIGURE 1