    User name paschke

    Log entry time 22:12:40 on October 6,2009

    Entry number 293541

    keyword=Compton: summary of coincidence runs

    
    Tried to establish coincidence runs with lower background.
    
    Electron detector position at 0mm.  
    
    
    ** Run #20161  100uA, ~10 minutes coincidences.  Looks good!
    
       http://www.jlab.org/~adaq/halog/html/0910_archive/091006183812.html
    
    ** Run #20162  100uA, ~1 hour coincidences.  Looks good.
    
    ** Run #20163  10uA.  0 coincidences found!
    
    
    
    ** Run #20164  - run failed, coda hung -  (Went to no prescale [100%] on photon 
    and electron, but also raised beam current... led to fifo overfill and hang.)
    
    ** Run #20165 short run, all prescale at 100 (means... take all).
    
    Scanned current to get a sense of this "threshold":
    
    ** Run #20166  50 uA, found coincidences. (Run 20167 same?) 
    
    ** Run #20168  30 uA, found coincidences. 
    
    ** Run #20169  20 uA, found coincidences.
    
    ** Run #20170  15 uA. Found coincidences.
    
    ** Run #20171  12-10 uA. Found coincidences at start, not after current went 
    from 12-10uA.
    
    ** Run #20172  10 uA.  No coincidences after 10 uA.
    
    
    Rebooted coda for no particular reason (getting ready for long run) and set 
    3-plane trigger to get bump free coincidences.  Should be a factor of 10 lower 
    statistics, right?
    
    ** Run #20173  15uA, 3-plane edet trigger. Rates unbelievably low, 
                much below 10x reduction in 2-plane edet runs. 
    
    ** Run #20174  15uA, 2-plane edet trigger.  Rates TOO LOW!!!!  WTF!?!
    
    ** Run #20176  same conditions, try again.  Need I say WTF?!?!
    
    Rather than raise beam current, went to 1-plane coincidence on edet.
    
    ** Run #20177  15uA, 1-plane edet trigger.  No coincidences at all.
    
    Tried to reboot coda, had some trouble getting back up.
    
    
    ** Run #20178  back to 100uA. 2 plane coincidence.  Seems fine.
    
    ** Run #20179 back to production (photon only) at 100 uA.  
    
    We'll return to get a low-current coincidence after being sure that the DAQ is 
    done acting dumb.