    User name Armstrong

    Log entry time 23:51:03 on October 24, 2009

    Entry number 296525

    keyword=Target studies... getting close to operating point

    Raster = 10mm x 10 mm Loop 1 Hydrogen

    Beam Current  HP Heater  Fan Speed  JT Valve   coolant  
     µA             (W)       (Hz)       (%)      (g/s)
    5              445         48         50         14
    20             323         48         50         14
    20             280         60         50         14       
    20             350         60         51         15.2
    20             515         60         57         18
    40             390         60         57         18
    60             230         60         57         18
    80              80         60         57         18
      Target seems not to be regulating here ; sits at 18.88 K, not 19.0 K. 
      Call Dave Meekins (again. sorry Dave...) He finds a problem in 
    the Beam Compensation software, and for now we have turned off beam
    compensation.
    

    80 140 60 57 18.2 80 215 60 58.6 19.2 98 85 60 58.6 19.2

    To try next: go up in Fan speed to 60 Hz or 72 Hz; may need to increase the JT valve as I expect the fan will heat the target up a bit. Watch Lumi widths and decide on optimal settings (perhaps reduce raster if possible)?